Impurity analyses of high-purity carbon monoxide gas using micro gas chromatography for development as a certified
Nobuhiro Matsumoto1, Takuro Watanabe, Kenji Kato
1National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST), 1-1-1 Umezono, Tsukuba, Ibaraki 305-8563, Japan. nobu-matsumoto@aist.go.jp
Abstract:
High-purity carbon monoxide (CO) gas as a certified reference material (CRM) was analyzed using a micro gas chromatograph (micro-GC) with a micro thermal conductivity detector. The main reason to select the micro-GC was its compact size to provide a safe experimental environment. Thus, both the CO gas cylinder and the micro-GC were placed inside a draft shield with a limited small space. Peaks of several impurities were found in chromatograms of the micro-GC. The main impurities in the CO gas were helium, hydrogen, oxygen, nitrogen and carbon dioxide. It was found that helium was the impurity of the highest concentration. High-accuracy calibration gas mixtures were used to construct calibration lines for the micro-GC. The mixtures were precisely prepared by the gravimetric blending method. Calibration lines had enough linearity and accuracy for quantitative analyses of the impurities. The values of detection limit of the impurities were 0.7-4μmol/mol. The purity of the high-purity CO gas was around 99.996%, which was the value estimated from the sum of the concentrations of the impurities.
More Related Videos
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...