C Alexopoulou1, E Kondili, M Plataki
1Department of Intensive Care Medicine, University Hospital of Heraklion, School of Medicine, University of Crete, Heraklion, Crete, Greece.
Objective:
To examine patient-ventilator asynchrony and sleep quality in non-sedated critically ill patients ventilated with proportional assist ventilation with load adjustable gain factors (PAV+) and pressure support (PSV).
Methods:
This was a randomized crossover physiological study conducted in an adult ICU at a tertiary hospital. Patients who exhibited patient-ventilator asynchrony on PSV were selected. Polysomnography was performed in these patients over 24 h, during which respiratory variables were continuously recorded. During the study period, each patient was randomized to receive alternating 4-h periods of PSV and PAV+ equally distributed during the day and night. Sleep architecture was analyzed manually using predetermined criteria. Patient-ventilator asynchrony was evaluated breath by breath using the flow-time and airway pressure-time waveforms.
Results:
Fourteen patients were studied. The majority (85.7 %) had either acute exacerbation of COPD as admission diagnosis or COPD as comorbidity. During sleep, compared to PSV, PAV+ significantly reduced the patient-ventilator asynchrony events per hour of sleep [5 (1-17) vs. 40 (4-443), p = 0.02, median (25-75th interquartile range)]. Compared to PSV, PAV+ was associated with slightly but significantly greater sleep fragmentation [18.8 (13.1-33.1) versus 18.1 (7.0-22.8) events/h, p = 0.01] and less REM sleep [0.0 % (0.0-8.4) vs. 5.8 % (0.0-21.9), p = 0.02).
Conclusions:
PAV+ failed to improve sleep in mechanically ventilated patients despite the fact that this mode was associated with better synchrony between the patient and ventilator. These results do not support the hypothesis that patient-ventilator synchrony plays a central role in determining sleep quality in this group of patients.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: