Sample Preparation for Probe Electrospray Ionization Mass Spectrometry
Published on: February 19, 2020
Electrospray ionization on porous spraying tips for direct sample analysis by mass spectrometry: enhanced detection
Melody Yee-Man Wong1, Ho-Wai Tang, Sin-Heng Man
1Department of Chemistry, Open Laboratory of Chemical Biology of the Institute of Molecular Technology for Drug Discovery and Synthesis, The University of Hong Kong, Pokfulam Road, Hong Kong SAR, PR China.
Rationale:
Despite various porous materials having been widely adopted as spraying tips for direct sample analysis using electrospray ionization mass spectrometry (ESI-MS), the effect of surface property and porosity of spraying tip materials on their analytical performances is not clear. Investigation of their relationships could provide insight into the proper choice and/or design of spraying tip materials for direct sample analysis.
Methods:
The effect of spraying tip materials with different polarities, including polyester and polyethylene (hydrophobic) and wood (hydrophilic), on the detection sensitivity for a variety of compounds, and on the ESI onset voltage, were studied using ESI-MS. The porosity of each type of spraying tip was characterized by scanning electron microscopy (SEM). Factors governing the detection sensitivity were determined based on the correlation of the detection sensitivity to the ESI onset voltage, the polarity, and the porosity of the spraying tip materials.
Results:
Hydrophobic tips (i.e., polyester and polyethylene) show better detection sensitivity for polar compounds but not for non-polar compounds, while hydrophilic tips (wooden tips) show the opposite effect. This phenomenon could be due to the difference in interaction between the analytes and the tips, causing the analytes to adsorb on the tip to different extents. In addition, the micro-porous nature of the tips could facilitate solvent diffusion for transporting analytes to the tip and maintain a stable spray for recording MS data. With the proper choice of spraying tip materials, trace amount of analytes at the picomole level can be detected with minimal sample pretreatment.
Conclusions:
Both the polarity and the porosity of the spraying tip materials could significantly affect detection sensitivity for a wide variety of analytes. With proper choice of spraying tip material, ESI on a porous spraying tip could be a sensitive method for the direct analysis of daily life samples.
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Chemical Ionization (CI) Mass Spectrometry
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
High-Performance Liquid Chromatography: Types of Detectors
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...