Raman spectroscopy for clinical-level detection of heparin in serum by partial least-squares analysis

Ali Momenpour T Monfared1, Vidhu S Tiwari, Markandey M Tripathi

  • 1University of Ottawa, School of Electrical Engineering and Computer Science, 800 King Edward, Ottawa, Ontario K1N 6N5, Canada.

Heparin is the most widely used anti-coagulant for the prevention of blood clots in patients undergoing certain types of surgeries including open heart surgeries and dialysis. The precise monitoring of heparin amount in patients' blood is crucial for reducing the morbidity and mortality in surgical environments. Based upon these considerations, we have used Raman spectroscopy in conjunction with partial least squares (PLS) analysis to measure heparin concentration at clinical level which is less than 10 United States Pharmacopeia (USP) in serum. The PLS calibration model was constructed from the Raman spectra of different concentrations of heparin in serum. It showed a high coefficient of determination (R2>0.91) between the spectral data and heparin level in serum along with a low root mean square error of prediction ~4 USP/ml. It enabled the detection of extremely low concentrations of heparin in serum (~8 USP/ml) as desirable in clinical environment. The proposed optical method has the potential of being implemented as the point-of-care testing procedure during surgeries, where the interest is to rapidly monitor low concentrations of heparin in patient's blood.

Related Concept Videos

Anticoagulant Drugs: Low-Molecular-Weight Heparins01:30

Anticoagulant Drugs: Low-Molecular-Weight Heparins

Hemostasis is a crucial process that prevents excessive blood loss from damaged blood vessels. It involves various mechanisms such as vasoconstriction, platelet adhesion and activation, and fibrin formation. The importance of each mechanism depends on the type of vessel injury. In contrast, thrombosis is the abnormal formation of a blood clot within the blood vessels, leading to potential complications if the clot obstructs blood flow. Thrombosis can be caused by increased coagulability of the...
Raman Spectroscopy Instrumentation: Overview01:26

Raman Spectroscopy Instrumentation: Overview

A conventional Raman spectrophotometer includes a laser source, a sample holding system, a wavelength selector, and a detector.
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview01:20

Raman Spectroscopy: Overview

The underlying principle of Raman spectroscopy is based on the interaction between light and matter, specifically molecules' inelastic scattering of photons. When a monochromatic beam of light, typically from a laser source, interacts with a sample, most scattered light has the same frequency as the incident light. This is known as Rayleigh scattering.
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...