A bench study of 2 ventilator circuits during helmet noninvasive ventilation
Giuliano Ferrone1, Flora Cipriani, Giorgia Spinazzola
1Department of Intensive Care and Anesthesia, Policlinico A Gemelli, Catholic University of Rome, Rome, Italy. giulianoferrone@yahoo.it
Objective:
To compare helmet noninvasive ventilation (NIV), in terms of patient-ventilator interaction and performance, using 2 different circuits for connection: a double tube circuit (with one inspiratory and one expiratory line) and a standard circuit (a Y-piece connected only to one side of the helmet, closing the other side).
Methods:
A manikin, connected to a test lung set at 2 breathing frequencies (20 and 30 breaths/min), was ventilated in pressure support ventilation (PSV) mode with 2 different settings, randomly applied, of the ratio of pressurization time to expiratory trigger time (T(press)/T(exp-trigger)) 50%/25%, default setting, and T(press)/T(exp-trigger) 80%/60%, fast setting, through a helmet. The helmet was connected to the ventilator randomly with the double and the standard circuit. We measured inspiratory trigger delay (T(insp-delay)), expiratory trigger delay (T(exp-delay)), T(press)), time of synchrony (T(synch)), trigger pressure drop, inspiratory pressure-time product (PTP), PTP at 300 ms and 500 ms, and PTP at 500 ms expressed as percentage of an ideal PTP500 (PTP500 index).
Results:
At both breathing frequencies and ventilator settings, helmet NIV with the double tube circuit showed better patient-ventilator interaction, with shorter T(insp-delay), T(exp-delay), and T(press); longer T(synch); and higher PTP300, PTP500, and PTP500 index (all P < .01).
Conclusions:
The double tube circuit had significantly better patient-ventilator interaction and a lower rate of wasted effort at 30 breaths/min.
More Related Videos
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: