Automated volumetric breast density estimation: a comparison with visual assessment
1Department of Radiology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Gangnam-gu, Seoul, Republic of Korea.
Clinical Radiology
|February 26, 2013
Summary
Automated volumetric breast density (VBD) measurement reliably assesses breast density, showing good agreement with radiologist visual assessment. Physical factors like x-ray tube target can influence VBD accuracy.
Area of Science:
- Radiology
- Medical Imaging
- Breast Health
Background:
- Breast density assessment is crucial for mammography interpretation and breast cancer risk stratification.
- Visual assessment of breast density is subjective and can vary between radiologists.
- Automated methods offer objective and reproducible breast density measurements.
Purpose of the Study:
- To compare automated volumetric breast density (VBD) measurements with radiologist visual assessment using BI-RADS categories.
- To identify factors influencing the agreement between automated VBD and visual assessment.
Main Methods:
- Retrospective analysis of 193 screening mammograms with negative findings.
- Qualitative BI-RADS density categorization by three radiologists.
- Automated measurement of VBD (% breast density) and volumetric density grade (VDG).
- Comparison of VDG with visual assessment and analysis of influencing physical factors.
Main Results:
- Good agreement was found between radiologist visual assessment and VDG (ICC = 0.757).
- VBD demonstrated a strong positive correlation with visual assessment (Spearman's ρ = 0.754, p < 0.001).
- Significant differences in VBD and x-ray tube target were observed between agreement and disagreement groups (p = 0.02 and p = 0.04).
Conclusions:
- Automated VBD measurement is a reliable and objective method for quantifying breast density.
- Physical factors, such as the x-ray tube target, can impact the agreement between automated VDG and visual density assessment.