Rupture of valsalva sinus after aortic root replacement with freestyle stentless bioprosthesis
Taichi Sakaguchi1, Shigeru Miyagawa, Hiroyuki Nishi
1Department of Cardiovascular Surgery, Osaka University Graduate School of Medicine, Osaka, Japan.
Abstract:
We report a case of aortic wall rupture in a patient with a Medtronic Freestyle stentless bioprosthesis (Medtronic Inc, Minneapolis, MN) after full root implantation. A 68-year-old man who underwent aortic root replacement with a Freestyle bioprosthesis 12 years prior was found to have a large pseudoaneurysm originating from the ruptured noncoronary porcine aortic sinus. A reoperation for aortic root replacement was successfully performed. Despite excellent durability of porcine aortic root bioprostheses, several cases with a ruptured aortic wall have been reported. Degeneration of the elastic tissue and inflammatory reactions may be the main mechanism of this potentially catastrophic complication.