Electromagnetic Navigation Transthoracic Nodule Localization for Minimally Invasive Thoracic Surgery
Published on: May 4, 2022
Rapid on-site evaluation and low registration error enhance the success of electromagnetic navigation bronchoscopy
Demet Karnak1, Aydin Ciledağ, Koray Ceyhan
1Department of Chest Diseases, Ankara University School of Medicine, Ankara, Turkey.
Background:
Electromagnetic navigation bronchoscopy (EMN) is a novel technology which allows localizing peripheral lung lesions and mediastinal lymph nodes for sampling and thus increasing diagnostic yield of Flexible Bronchoscopy.
Objectives:
A prospective study was conducted to investigate the diagnostic yield of EMN with lower average fiducial target registration error (AFTRE) and rapid on-site evaluation (ROSE).
Methods:
Consecutive patients with peripheral lung lesion (PL) or enlarged mediastinal lymph node (MLN) which could not be diagnosed by conventional techniques and/or if the patients were not suitable for such interventions were included. The navigation procedure was continued once registration error was reached below/equal to the absolute value of 5 mm. ROSE was performed by an expert cytopathologist.
Results:
A total of 76 patients; 22 having only PLs, 41 having only MLNs, and 13 having both PLs and MLNs together were enrolled. Thirty-two of 35 PLs (91.4%) and 85 of 102 MLNs (83.3%) were successfully sampled. Overall diagnostic yield was 89.5%. PLs and MLNs were further grouped according to their size (PLs: <20 mm vs ≥20 mm, MLNs: <15 mm vs ≥15 mm). The sampling yield was independent of size for both PL and MLN (P = 1.00, P = 0.38). In diagnostic EMN cases, mean AFTRE was 4.33 ± 0.71 mm, whereas it was 5.16 ± 0.05 mm (P = 0.008) in nondiagnostics. The total duration of procedure was 36.17 ± 9.13 min. Pneumothorax was observed in three patients (3.9%).
Conclusion:
EMN with low AFTRE in combination with ROSE is a reliable method with high sampling and/or diagnostic rate in PLs and MLNs.