Measurement of phantomless thoracic bone mineral density on coronary artery calcium CT scans acquired with various CT
Matthew J Budoff1, Jennifer M Malpeso, Irfan Zeb
1Los Angeles Biomedical Research Institute, Harbor-UCLA Medical Center, 1124 W Carson St, Torrance, CA 90502-2006, USA. mbudoff@labiomed.org
Purpose:
To determine the accuracy and precision of thoracic phantomless bone mineral density (BMD) measurements obtained on coronary artery calcium (CAC) computed tomography (CT) scans by using a variety of commercially available CT scanners.
Materials And Methods:
The institutional review board approved this Health Insurance Portability and Accountability Act-compliant study. A total of 4126 asymptomatic subjects (2022 [49%] men, 2104 [51%] women; mean age, 63.7 years ± 11.8 [standard deviation]) underwent CAC CT with the use of a quantitative CT calibration phantom for evaluation of subclinical atherosclerosis. Two hundred eighty subjects also underwent CT of the chest, abdomen, and pelvis (C7 through L5). Mean BMD of three consecutive thoracic vertebrae (in the T7-T10 range) was measured in all 4126 subjects. Individual calibration factors for each phantom insert and a general calibration factor for the spine were determined for each CT scanner model. The study population was then divided into three subgroups: All calibration factors were generated from group 1 (n = 1536) and were applied and tested in group 2 (n = 1587), and effects of various image acquisition parameters were assessed in group 3 (n = 1003). Accuracy (bias) and precision of thoracic phantomless BMD measurements across 14 CT scanner models from five manufacturers were determined.
Results:
Phantomless BMD values correlated highly with standard phantom-based quantitative CT BMD values (r = 0.987, P < .001). Bias was 3.9% ± 1.4 for phantomless BMD measurements, and the mean coefficient of variation for the general calibration factor was 4.9% ± 2.4.
Conclusion:
Phantomless BMD can be measured accurately on CAC CT scans acquired with a variety of CT scanners without additional radiation exposure.
More Related Videos
06:57Semi-Automatic Graphical Tool for Measuring Coronary Artery Spatially Weighted Calcium Score from Gated Cardiac Computed Tomography Images
Published on: September 22, 2023
08:02Novel Quantification Protocol for Cardiovascular Calcification Progression Using Longitudinal MicroPET/MicroCT Images
Published on: November 15, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...