Incorrect installation of free-flow clamp in B. Braun Infusomat Space infusion pump may result in gravity flow
Abstract:
Gravity flow--potentially resulting in a dangerous overinfusion--may occur with the B. Braun Infusomat Space infusion pump if the infusion set's free-flow clamp is not correctly inserted in the pump. The manufacturer is currently evaluating measures to reduce the likelihood of this error. In the interim, facilities should ensure that users are familiar with proper use of the pump and are trained to recognize signs that the infusion set was loaded incorrectly.