Quality control lockout function on Medtronic ACT Plus coagulation timer system can be bypassed for same-patient
Health Devices
|March 1, 2013
Summary
The Medtronic ACT Plus analyzer may not lock out testing after quality control (QC) is due. This can lead to inaccurate results and dangerous heparin therapy decisions.
Area of Science:
- Medical Device Technology
- Clinical Pathology
- Laboratory Medicine
Background:
- The Medtronic ACT Plus automated coagulation timer system is used for activated clotting time (ACT) analysis.
- Quality control (QC) is essential for ensuring the accuracy of diagnostic test results.
Purpose of the Study:
- To identify a potential malfunction in the Medtronic ACT Plus system's QC lockout function.
- To alert users to a critical issue that may compromise patient safety.
Main Methods:
- Investigated the behavior of the Medtronic ACT Plus system when QC has not been performed.
- Observed the system's response to sample analysis requests following overdue QC intervals.
Main Results:
- The ACT Plus system's QC lockout function may fail to activate on single-patient use.
- The device provides an alert (beeping and on-screen message) but proceeds with sample analysis.
- This can result in the reporting of inaccurate ACT results.
Conclusions:
- Inaccurate ACT results can lead to incorrect and potentially dangerous heparin therapy decisions.
- Users must be aware that the device may not enforce QC checks.
- It is crucial for users to manually verify QC completion before analyzing patient samples.