Trending of impedance-monitored cardiac variables: method and statistical power analysis of 100 control studies in a

J J Mickell1, S E Lucking, F C Chaten

  • 1Department of Pediatrics, Children's Medical Center, Medical College of Virginia, Virginia Commonwealth University, Richmond.

Insights

The NCCOM3-R6 monitor accurately tracks cardiac output and other cardiovascular metrics in pediatric patients. Large sample sizes enable detection of significant clinical changes, ensuring reliable patient monitoring.

Area of Science:

  • Cardiovascular Physiology
  • Critical Care Medicine
  • Biomedical Engineering

Background:

  • Continuous hemodynamic monitoring is crucial in pediatric intensive care units (PICUs).
  • Thoracic electrical bioimpedance (TEB) offers a non-invasive method for assessing cardiovascular function.
  • The NCCOM3-R6 device utilizes TEB to measure cardiac output and other vital parameters.

Purpose of the Study:

  • To evaluate the reliability and sensitivity of the NCCOM3-R6 monitor for continuous cardiovascular assessment in pediatric patients.
  • To determine the feasibility of detecting clinically significant changes in cardiovascular variables using TEB monitoring in a large cohort.
  • To establish the required sample size for detecting specific changes in cardiac output with adequate statistical power.

Main Methods:

  • Data from 100 control studies involving 40 pediatric ICU patients (age 0.04-20.39 years) were analyzed.
  • Cardiac output and five other cardiovascular variables were continuously monitored using the NCCOM3-R6 device.
  • Data were averaged over 5-minute intervals for 130 minutes, with analysis of individual and aggregated study results.

Main Results:

  • Individual 5-minute averages of cardiac output showed 99% agreement within +/- 44% of the baseline for each study.
  • Aggregated data from 100 studies revealed no significant deviation from baseline for any measured variable over a 2-hour period (p > 0.70).
  • A sample size of 100 studies allowed detection of a +/- 5% change in cardiac output with 95% power at p < 0.005.

Conclusions:

  • The NCCOM3-R6 monitor provides reliable cardiovascular variable measurements in pediatric ICU patients.
  • Sufficiently large sample sizes are essential for detecting subtle yet clinically significant hemodynamic changes using TEB.
  • The NCCOM3-R6, with adequate sample size, can effectively identify pharmacologic or procedural impacts on cardiovascular status in critically ill children.