Corneal thickness and volume measurements by swept source anterior segment optical coherence tomography in normal
Reina Fukuda1, Tomohiko Usui, Takashi Miyai
1Department of Ophthalmology, University of Tokyo School of Medicine, Tokyo, Japan.
Purpose:
To evaluate central corneal thickness (CCT) and corneal volume (CV) in healthy eyes using swept source anterior segment optical coherence tomography (SS-AS-OCT) and other devices.
Methods:
Thirty-three healthy right eyes of 33 subjects were investigated. CCT was measured with SS-AS-OCT, rotating Scheimpflug camera, ultrasonic pachymetry and specular microscopy. CV was determined within a diameter of 10 mm at the center using SS-AS-OCT and Scheimpflug camera.
Results:
Mean CCT was 523.5 ± 25.2 μm by SS-AS-OCT, 523.9 ± 26.1 μm by Scheimpflug camera, 532.1 ± 26.6 μm by ultrasonic pachymetry, and 525.5 ± 33.5 μm by specular microscopy. The CCTs measured with SS-AS-OCT, Scheimpflug camera and specular microscopy were significantly thinner than those measured with ultrasonic pachymeter (p < 0.001). The mean CV within a diameter of 10 mm at the center was 57.2 ± 3.0 mm(3) by SS-AS-OCT and 59.4 ± 2.9 mm(3) by Scheimpflug camera. Scheimpflug measurements of CV were statistically larger than SS-AS-OCT measurements (p = 0.0008). Statistically significant correlation was found between the CCT and CV measurements of each imaging devices (p < 0.0001, r = 0.527, and p < 0.0001, r = 0.749, respectively).
Conclusion:
SS-AS-OCT enables the measurements of CCT and CV, demonstrating agreement with other devices. CV in addition to CCT measurement may serve as a practical parameter of the corneal endothelial pump function.