First experience with monochromatic coronary computed tomography angiography from a 64-slice CT scanner with Gemstone
Tobias A Fuchs1, Julia Stehli, Michael Fiechter
1Department of Radiology, Cardiac Imaging, University Hospital Zurich, Ramistrasse 100, NUK C 42, CH-8091 Zurich, Switzerland.
Background:
New technology combining dual-energy CT with the latest gemstone detectors for spectral imaging (GSI) can be used to synthesize monochromatic images that mimic images as if different monochromatic x-ray sources were used.
Objective:
The aim of the study was to evaluate the optimal combination of monochromatic image energy and adaptive statistical iterative reconstruction (ASiR) for monochromatic reconstruction of coronary CT angiography (CTA) images with the use of GSI.
Methods:
Twenty consecutive patients underwent coronary CTA on a GSI capable 64-slice CT scanner (Discovery CT 750 High Definition, GE Healthcare). In 7 sets of monochromatic images (60, 65, 70, 75, 80, 90, and 110 keV; each with increasing contributions of ASiR, ie, 0%, 20%, 40%, 60%, and 80%; n = 35 reconstructions per patient), signal-to-noise (aortic root) and contrast-to-noise (left main artery) ratios were assessed. Signal-to-noise ratio, contrast-to-noise ratio, and image quality (graded on a 5-point Likert scale) were assessed in all above monochromatic reconstructions and compared with the respective standard (conventional polychromatic) image.
Results:
Compared with conventional polychromatic images, reconstructions with 60 keV and 80% ASiR showed the highest improvement in contrast-to-noise (144%; P < 0.001) and signal-to-noise ratio (173%; P < 0.001). Image quality reached a plateau at 65-75 keV with 40%-60% ASiR blending, yielding a maximal image quality score improvement of 50% compared with conventional imaging (P < 0.001).
Conclusion:
In coronary CTA with low radiation technique (mean radiation dose, 1.8 ± 0.7 mSv), GSI with monochromatic reconstructions (65-75 keV) and ASiR (40%-60%) offers significant noise reduction and image quality improvement.
More Related Videos
06:57Semi-Automatic Graphical Tool for Measuring Coronary Artery Spatially Weighted Calcium Score from Gated Cardiac Computed Tomography Images
Published on: September 22, 2023
09:09Single Nuclei Isolation from Coronary Endarterectomy Tissue of Coronary Artery Bypass Graft Patients
Published on: April 3, 2026
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System IV: CMRI