Validation of an automated algorithm for detecting apneas and hypopneas by acoustic analysis of breath sounds
Related Experiment Video
Updated: May 13, 2026

Asthma Detection Research Based on Voice Signal Processing and Machine Learning
Published on: July 22, 2025
Hisham Alshaer1, Geoff R Fernie, Ellen Maki
1University Health Network Toronto Rehabilitation Institute, iDAPT - Intelligent Design for Adaptation, Participation and Technology, Canada. hisham.alshaer@uhn.ca
Background:
Sleep-disordered breathing (SDB) is common and is associated with increased risk for cardiovascular disease. However, most patients remain undiagnosed due to lack of access to sleep laboratories. We therefore tested the validity of a single-channel monitoring setup that captures and analyzes breath sounds (BSs) to detect SDB.
Methods:
BS were recorded from 50 patients undergoing simultaneous polysomnography (PSG). Using custom-designed automatic software, BS were subjected to a set of pattern recognition rules to identify apneas and hypopneas from which the acoustic apnea-hypopnea index (AHI-a) was calculated. Apneas and hypopneas from PSG were scored blindly by three technicians according to two criteria; one relying solely on the drop of the respiratory signal by >90% for an apnea and by 50% to 90% for a hypopnea (TV50 criteria), and another that also required a desaturation or an arousal for a hypopnea (American Association of Sleep Medicine [AASM] criteria). PSG AHI (AHI-p) was calculated for each technician according to both criteria.
Results:
There was no significant difference between AHI-p scores according to TV50 and AASM criteria. AHI-a was strongly correlated with AHI-p according to both TV50 (R=94%) and AASM criteria (R=93%). Bland-Altman plot analysis revealed that 98% and 92% of AHI-a fell within the limits of agreement for AHI-p according to TV50 and AASM criteria, respectively. Based on a diagnostic cutoff of AHI-p≥10 for SDB, overall accuracy of AHI-a reached 88% and negative predictive value reached 100%.
Conclusion:
Acoustic analysis of BS is a reliable method for quantifying AHI and diagnosing SDB compared to simultaneous PSG.
Related Concept Videos
Physical Assessment of the Respiratory Tract IV: Auscultation
Breath Sounds
Breath sounds are categorized into vesicular, bronchovesicular, and bronchial.
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: