Related Experiment Video
Updated: May 13, 2026

Full-Field Optical Coherence Microscopy for Histology-Like Analysis of Stromal Features in Corneal Grafts
Published on: October 21, 2022
Open-access operating algorithms for commercial videokeratographer and improvement of corneal sampling
Julián Espinosa1, David Mas, Jorge Pérez
1Instituto Universitario de Física Aplicada a las Ciencias y las Tecnologías, Universidad de Alicante, Alicante, Spain. julian.espinosa@ua.es
Abstract:
We present an algorithm to process images of reflected Placido rings captured by a commercial videokeratoscope. Raw data are obtained with no Cartesian-to-polar-coordinate conversion, thus avoiding interpolation and associated numerical artifacts. The method provides a characteristic equation for the device and is able to process around 6 times more corneal data than the commercial software. Our proposal allows complete control over the whole process from the capture of corneal images until the computation of curvature radii.