Respiration based steering for high intensity focused ultrasound liver ablation
Andrew B Holbrook1, Pejman Ghanouni, Juan M Santos
1Department of Radiology, Stanford University, Stanford, California, USA.
Purpose:
Respiratory motion makes hepatic ablation using high intensity focused ultrasound (HIFO) challenging. Previous HIFU liver treatment had required apnea induced during general anesthesia. We describe and test a system that allows treatment of the liver in the presence of breathing motion.
Methods:
Mapping a signal from an external respiratory bellow to treatment locations within the liver allows the ultrasound transducer to be steered in real time to the target location. Using a moving phantom, three metrics were used to compare static, steered, and unsteered sonications: the area of sonications once a temperature rise of 15°C was achieved, the energy deposition required to reach that temperature, and the average rate of temperature rise during the first 10 s of sonication. Steered HIFU in vivo ablations of the porcine liver were also performed and compared to breath-hold ablations.
Results:
For the last phantom metric, all groups were found to be statistically significantly different (P ≤ 0.003). However, in the other two metrics, the static and unsteered sonications were not statistically different (P > 0.9999). Steered in vivo HIFU ablations were not statistically significantly different from ablations during breath-holding.
Conclusions:
A system for performing HIFU steering during ablation of the liver with breathing motion is presented and shown to achieve results equivalent to ablation performed with breath-holding.