Related Experiment Video
Updated: May 13, 2026

Multi-enzyme Screening Using a High-throughput Genetic Enzyme Screening System
Published on: August 8, 2016
A 96-well electrochemical method for the screening of enzymatic activities
Sofiène Abdellaoui1, Alexandre Noiriel, Robert Henkens
1GEMBAS (Génie Enzymatique, Membranes Biomimétiques et Assemblages Supramoléculaires), ICBMS UMR 5246, Université Lyon 1, CNRS, INSA Lyon, Villeurbanne, France.
Abstract:
The rapid electrochemical screening of enzyme activities in bioelectronics is still a challenging issue. In order to solve this problem, we propose to use a 96-well electrochemical assay. This system is composed of 96 screen-printed electrodes on a printed circuit board adapted from a commercial system (carbon is used as the working electrode and silver chloride as the counter/reference electrode). The associated device allows for the measurements on the 96 electrodes to be performed within a few seconds. In this work, we demonstrate the validity of the screening method with the commercial laccase from the fungus Trametes versicolor. The signal-to-noise ratio (S/N) is found to be the best way to analyze the electrochemical signals. The S/N follows a saturation-like mechanism with a dynamic linear range of two decades ranging from 0.5 to 75 ng of laccase (corresponding to enzymatic activities from 62 × 10(-6) to 9.37 × 10(-3) μmol min(-1)) and a sensitivity of 3027 μg(-1) at +100 mV versus Ag/AgCl. Laccase inhibitors (azide and fluoride anions), pH optima, and interfering molecules could also be identified within a few minutes.

