Quantification and reduction of reflux during embolotherapy using an antireflux catheter and tantalum microspheres:
Aravind Arepally1, James Chomas, Dara Kraitchman
1Division of Interventional Radiology, Piedmont Radiology, 1984 Peachtree Rd., Suite 505, Atlanta, GA 30309, USA. aarepal@gmail.com
Purpose:
To demonstrate in a porcine model that reflux during embolotherapy can be relatively quantified (ie, as embolization efficiency) and that nontarget embolization can be eliminated by using an antireflux microcatheter.
Materials And Methods:
Renal artery embolization was performed with radiopaque tantalum microspheres (concentration of 1 g/20 mL) in three swine. Second-order right renal arteries (n = 3) underwent embolization with a 3-F antireflux catheter, and second-order left renal arteries (n = 3) underwent embolization with a 4-F end-hole catheter as a control. After embolization, kidneys were explanted and underwent micro-computed tomographic (microCT) imaging. Three-dimensional volumetric and multiplanar imaging of the kidneys was performed to assess vascular distribution. Digital Imaging and Communications in Medicine data were analyzed, with a threshold algorithm used to create binary images. The number of positive values in a region of interest in the target embolized tissue (upper pole or lower pole) and the nontarget adjacent tissue was determined, and embolization efficiency was calculated. Wilcoxon rank-sum statistical analysis was performed to compare nontarget embolization between infusion catheters.
Results:
All renal arteries underwent successful embolization with tantalum microspheres, with 20 mL (1 g) administered in all dose deliveries. MicroCT provided high-resolution visualization of the renal parenchyma at 70-μm resolution. In control renal arteries, a standard 4-F end-hole catheter had an embolization efficiency of 72%± 13. In experimental renal arteries, the antireflux microcatheter had an embolization efficiency of 99.9%± 1.0 (P< . 05).
Conclusions:
A significant decrease in nontarget embolization (ie, reduction in reflux) was possible with an antireflux microcatheter compared with a conventional end-hole catheter.