Cross-camera comparison of ROI-based semi-quantitative ¹²³I-IBZM SPECT data in healthy volunteers using an
Susanna Jakobson Mo1, Anne Larsson2, Lennart Johansson2
1Department of Radiation Sciences, Diagnostic Radiology, Umeå University, Umeå susanna.jakobson.mo@diagrad.umu.se.
Background:
In (123)I-Iolopride (IBZM) SPECT reference values may diverge between camera systems. If multicenter pooling of normal material databases is needed, differences in measured semi-quantitative data due to equipment performance and reconstruction parameters have to be investigated in each instance to determine the comparability.
Purpose:
To explore the differences in (123)I-IBZM measured uptake ratios between two different gamma cameras in healthy controls, the intra-rater reproducibility of the image evaluation method and the possibility to equalize uptake ratios by calibration through an anthropomorphic phantom.
Material And Methods:
Differences in ROI-based semi-quantitative data from two different gamma camera systems, the three-headed brain dedicated Neurocam and the two-headed multipurpose hybrid system Infinia Hawkeye, were studied using image data from a group of healthy volunteers and an anthropomorphic brain-phantom scanned with both cameras. Several reconstruction methods and corrections were applied. To test the reliability of the ROI method, the intra-observer reproducibility was determined for the ROI method in this study.
Results:
The ROI method had a high reliability. Differences in mean measured uptake (123)I-IBZM ratios in healthy controls varied between 2.9% and 6.5% depending on reconstruction and correction for attenuation and scatter. After calibration, the differences decreased. There were no statistically significant differences between corrected ratios from the two camera systems in the study when images were reconstructed with attenuation correction.
Conclusion:
The conformity of uptake ratios in attenuation corrected (123)I-IBZM images derived from the two different cameras was improved by using an anthropomorphic phantom for calibration.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography