Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
[The image noise effect on the results of Gamma knife dosimetry parameters test]
Xiaojun Cheng1, Conghua Zhang, Chuanpeng Hu
1Radiation Protection Department, Henan Institute of Occupational Medicine, Zhengzhou 450052, China. hnfsws@163.com
Abstract:
In order to analyze the image noise effect on the results of Gamma knife dosimetry parameter test, we tested the dosimetry parameters of the Gamma knives according to GBZ 168-2005. Radiological protection standards of X (gamma)-ray stereotactic radiosurgery for head treatment. Dose analysis software was applied to examine the testing film before and after image denoising, and SPSS 11.0 software was used for statistical analysis. The results showed that there was a significant difference in the results of the maximum deviation between radiation field size and its nominal value (t = 7.600, P < 0.01) and the radiation field's penumbra region width of collimators also had significantly different sizes (t = 5.334, P < 0.01) before and after image denoising. This study indicated that the image noise could influence the results of testing Gamma knife dosimetry parameters, so as to cause deviations.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET