A four-pixel matched collimator for high-sensitivity SPECT imaging
Atsuro Suzuki1, Wataru Takeuchi, Takafumi Ishitsu
1Central Research Laboratory, Hitachi Ltd, Hitachi-shi, Ibaraki-ken, Japan. atsuro.suzuki.rc@hitachi.com
Abstract:
We propose a wide aperture parallel-hole collimator that we call a 4-pixel matched collimator (4-PMC) for high-sensitivity SPECT imaging. The hole size of the 4-PMC is matched to four detector pixels; that is, there are four (2 × 2) pixels per collimator hole. By contrast, a 1-pixel matched collimator (1-PMC) is defined as a collimator whose hole size is matched to one detector pixel. We evaluated four types of collimator (high-resolution collimator versions and high-sensitivity collimator versions of both 4-PMC and 1-PMC) by simulation. SPECT images of a cylindrical phantom with cold spots in the noise-free condition demonstrated that the 4-PMC provided a higher-contrast image than the 1-PMC for the same collimator version. In addition, SPECT images at the noise level corresponding to a human cerebral blood flow study suggested that the high-sensitivity version of the 4-PMC provided the highest contrast image among the four collimator types. In conclusion, the high-sensitivity SPECT system using the 4-PMC can improve the trade-off between spatial resolution and sensitivity and will consequently provide improved image contrast for clinical studies of the human brain compared with the SPECT system using the 1-PMC.