An HPLC-DAD method for the simultaneous quantification of opicapone (BIA 9-1067) and its active metabolite in human
Daniela Gonçalves1, Gilberto Alves, Ana Fortuna
1Laboratory of Pharmacology, Faculty of Pharmacy, University of Coimbra, Pólo das Ciências da Saúde, Azinhaga de Santa Comba, 3000-548 Coimbra, Portugal.
Abstract:
Opicapone (BIA 9-1067) is a novel catechol-O-methyltransferase inhibitor presently under clinical development as an adjuvant in the pharmacotherapy of Parkinson's disease. This report describes the development and validation of a bioanalytical assay for the simultaneous quantification of opicapone and its active metabolite (BIA 9-1079) in human plasma. The method herein reported is based on high-performance liquid chromatography coupled with diode-array detection (HPLC-DAD) and the sample preparation consists of a plasma protein precipitation step followed by liquid-liquid extraction. Chromatographic separation of the analytes (opicapone and BIA 9-1079) and the internal standard (tamoxifen) was achieved in less than 10 min on a reversed-phase C18 column at 25 °C by applying a gradient elution program using a mobile phase composed of 0.05 M monosodium phosphate solution adjusted to pH 2.45 (A) and acetonitrile (B) pumped at 0.8 mL min(-1). Opicapone and the internal standard were monitored at 271 nm while BIA 9-1079 was assessed at 257 nm. Calibration curves of both analytes were linear (r(2) ≥ 0.997) in the concentration range of 25-3000 ng mL(-1) and their limits of quantification were established to be 25 ng mL(-1). The overall precision did not exceed 13.2% and the accuracy was within ±11.1%. Several drugs potentially co-administered with opicapone were tested and they did not interfere at the retention times of the analytes (opicapone and BIA 9-1079) and internal standard. The method was then successfully applied for quantifying opicapone and its active metabolite (BIA 9-1079) in plasma samples obtained from a healthy subject enrolled in a clinical trial.
More Related Videos
Related Concept Videos
Drug Concentrations: Measurements
Plasma —...
Measurement of Bioavailability: Pharmacodynamic Methods