Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
MR-driven metal artifact reduction in PET/CT
G Delso1, S Wollenweber, A Lonn
1GE Healthcare, Waukesha, WI, USA. gaspar.delso@usz.ch
Abstract:
Among the proposed system architectures capable of delivering positron emission tomography/magnetic resonance (PET/MR) datasets, tri-modality systems open an interesting field in which the synergies between these modalities can be exploited to address some of the problems encountered in standalone systems. In this paper we present a feasibility study of the correction of dental streak artifacts in computed tomography (CT)-based attenuation correction images using complementary MR data. The frequency and severity of metal artifacts in oncology patients was studied by inspecting the CT scans of 152 patients examined at our hospital. A prospective correction algorithm using CT and MR information to automatically locate and edit the region affected by metal artifacts was developed and tested retrospectively on data from 15 oncology patients referred for a PET/CT scan. In datasets without malignancies, the activity in Waldeyer's ring was used to measure the maximum uptake variation when the proposed correction was applied. The measured bias ranged from 10% to 30%. In datasets with malignancies on the slices affected by artifacts, the correction led to lesion uptake variations of 6.1% for a lesion 3 cm away from the implant, 1.5% for a lesion 7 cm away and <1% for a lesion 8 cm away.