Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
MR-driven metal artifact reduction in PET/CT.
G Delso1, S Wollenweber, A Lonn
1GE Healthcare, Waukesha, WI, USA. gaspar.delso@usz.ch
Physics in Medicine and Biology
|March 13, 2013
Summary
This study presents a new method to correct dental artifacts in CT scans using MRI data. This technique improves image quality for oncology patients undergoing PET/CT scans.
Area of Science:
- Medical Imaging
- Radiology
- Biomedical Engineering
Background:
- Tri-modality systems combining PET/MR offer synergistic advantages over standalone systems.
- Metal artifacts, particularly dental streak artifacts, pose challenges in CT-based attenuation correction for PET/CT.
- Assessing the frequency and severity of these artifacts is crucial for accurate imaging in oncology patients.
Purpose of the Study:
- To investigate the feasibility of correcting dental streak artifacts in CT images using complementary MR data.
- To develop and test an algorithm for automatic localization and editing of metal artifact regions.
- To evaluate the impact of artifact correction on radiotracer uptake measurements in PET/CT scans.
Main Methods:
- A prospective correction algorithm was developed, utilizing both CT and MR information.
- The algorithm was retrospectively tested on PET/CT data from 15 oncology patients with metal artifacts.
- Artifact correction efficacy was assessed by measuring uptake variations in specific regions (Waldeyer's ring) and in lesions near implants.
Main Results:
- The developed algorithm successfully located and edited regions affected by metal artifacts.
- In non-malignant datasets, the correction reduced bias, with measured variations ranging from 10% to 30%.
- In datasets with malignancies, artifact correction resulted in minimal lesion uptake variations (<1% to 6.1% depending on distance from the implant).
Conclusions:
- Complementary MR data can effectively correct dental streak artifacts in CT attenuation correction images.
- The proposed correction algorithm demonstrates feasibility and minimizes the impact of artifacts on quantitative PET measurements.
- This technique holds promise for improving the accuracy of PET/CT imaging in oncology patients with dental implants.