A Simple Approach to Perform TEER Measurements Using a Self-Made Volt-Amperemeter with Programmable Output Frequency
Published on: October 5, 2019
A four-electrode low frequency impedance spectroscopy measurement system using the AD5933 measurement chip
1LPEM, UMR8213, ESPCI-ParisTech, 10 Rue Vauquelin, 75005 Paris, France. cedric.margo@espci.fr
Abstract:
This paper presents the design of a four electrode impedance measurement circuit dedicated to bioimpedance embedded applications. It extends the use of the AD5933 measurement chip to allow it to operate in a four electrode configuration in order to limit the effects of the parasitic impedances between the medium under test and the electrodes. The circuit has shown a good measurement accuracy on various test circuits. In association with a four microband electrode system it has been successfully used to characterize small physiological samples (50 μl) with conductivities ranging from 0.14 to 1.2 S m(-1). It can be used as an alternative bioimpedance measurement approach for embedded applications operating in the four electrode configuration.