Agreement between spirometers: a challenge in the follow-up of patients and populations?
M W Gerbase1, E Dupuis-Lozeron, C Schindler
1Division of Pulmonary Medicine, University Hospitals of Geneva, Geneva, Switzerland. margaret.gerbase@hcuge.ch
Background:
Long-term cohort studies and lung function laboratories are confronted with the need for replacement of spirometers. Lack of agreement between spirometers might affect the longitudinal comparison of data, notably when replacing conventional by portable spirometers.
Objectives:
To compare the handheld EasyOne (EO) with the conventional SensorMedics (SM) spirometer, and to analyze the interdevice reproducibility of EO spirometers.
Methods:
In total, 82 volunteers completed spirometry sessions with 1 SM and 2 of 3 EO spirometers following a Latin square design. Analyses of differences in forced vital capacity (FVC), forced expiratory flow in 1 s (FEV1), FEV1/FVC and mean forced expiratory flow calculated between 25 and 75% of the FVC between spirometers used a mixed effect model with a random intercept for each subject and the effect of the device as fixed effect adjusted for sex, age, height and order of spirometer tested. Bland-Altman plots show the 95% limits of agreement.
Results:
Comparisons between EO and SM showed relatively small mean differences of <3%, but systematically lower values for FVC and FEV1 in all EO devices. The 95% agreement exceeded the limits for FEV1 by 50 ml in 2 EO spirometers. The EO interdevice comparisons showed mean differences and limits of agreement within established thresholds, thus indicating fair accuracy when comparing devices. Repeats with the same spirometer did not result in statistically significant differences.
Conclusions:
This study suggests fair agreement between the handheld and the conventional spirometer. Differences slightly exceeding limits for FEV1 in 2 EO devices might be considered mostly irrelevant for clinical practice. However, the systematically lower FVC and FEV1 observed with EO may be significant for epidemiological studies, thus justifying inspection before replacing devices.
Related Concept Videos
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Physical Assessment of the Respiratory Tract I: Health History
Subjective Data
Subjective data provides vital information about the patient's health history and symptoms. This data is typically collected through interviews in which patients describe their experiences, symptoms, and concerns.
Health history and key...
Chronic Obstructive Pulmonary Disease-I: Introduction
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...