Non-invasive continuous arterial pressure measurement based on radial artery tonometry in the intensive care unit: a
B Saugel1, A S Meidert, A Hapfelmeier
1II. Medizinische Klinik und Poliklinik, Klinikum rechts der Isar der Technischen Universität München, Ismaninger Strasse 22, D-81675 München, Germany. bernd.saugel@gmx.de
Background:
The T-Line TL-200pro (TL-200pro) device (Tensys Medical, Inc., San Diego, CA, USA), based on radial artery tonometry, provides an arterial pressure (AP) waveform and beat-to-beat values of systolic arterial pressure (SAP), mean arterial pressure (MAP), and diastolic arterial pressure (DAP). The aim of the study was to evaluate this non-invasive technique for continuous AP monitoring in medical intensive care unit (ICU) patients.
Methods:
Arterial pressure measurements obtained using the TL-200pro technology were compared using Bland-Altman analysis with values measured directly from a femoral arterial catheter in 34 ICU patients.
Results:
Arterial pressure values were analysed and compared in 4502 averaged 10-beat epochs. A bias of +0.72 mm Hg (95% limits of agreement -9.37 to +10.82 mm Hg) was observed for MAP. For SAP and DAP, there was a mean difference of -1.39 mm Hg (95% limits of agreement -18.74 to +15.96 mm Hg) and +4.36 mm Hg (95% limits of agreement -8.66 to +17.38 mm Hg), respectively. The percentage error for MAP, SAP, and DAP was 12%, 14%, and 21%, respectively.
Conclusions:
Arterial pressure measurement based on radial artery tonometry using the TL-200pro technology is feasible in medical ICU patients. The TL-200pro system is capable of providing MAP values with high accuracy (low mean difference) and precision (narrow limits of agreement) compared with MAP measured invasively using a femoral arterial catheter. The TL-200pro technology is promising for the measurement of SAP and DAP but further development is necessary to improve accuracy and precision.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: