Temporal instabilities associated with a planar high purity germanium detector.

N D Pugh1, G M Owen, W D Evans

  • 1Department of Medical Physics and Bioengineering, University Hospital of Wales, Cardiff, UK.

Physics in Medicine and Biology
|May 1, 1990
PubMed
Summary

Grooved germanium detectors show time-varying counts due to charge collection issues. This impacts x-ray fluorescence sensitivity, but collimation can mitigate the problem.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Social Norms Change and Cigarette Smoking: A Longitudinal Study in Mainland China.

Nicotine & tobacco research : official journal of the Society for Research on Nicotine and Tobacco·2026
Same author

Social Media-Based Interventions for Health Behavior Change in Low- and Middle-Income Countries: Systematic Review.

Journal of medical Internet research·2022
Same author

Design of a novel digital intervention to promote healthy weight management among postpartum African American women.

Contemporary clinical trials communications·2019
Same author

Development of the place-based Adelante social marketing campaign for prevention of substance use, sexual risk and violence among Latino immigrant youth.

Health education research·2018
Same author

Central arterial stiffness and diastolic dysfunction are associated with insulin resistance and abdominal obesity in young women but polycystic ovary syndrome does not confer additional risk.

Human reproduction (Oxford, England)·2014
Same author

Quantification of the activity of tritium produced during the routine synthesis of (18)F fluorodeoxyglucose for positron emission tomography.

Journal of radiological protection : official journal of the Society for Radiological Protection·2014

Area of Science:

  • Semiconductor detector physics
  • Materials science

Background:

  • High purity germanium (HPGe) detectors are crucial for precise elemental analysis.
  • Grooved detector geometries are employed to enhance performance but can introduce complexities.

Purpose of the Study:

  • Investigate the origin of time-variant counting rates observed in grooved planar HPGe detectors.
  • Analyze the impact of these count rate variations on x-ray fluorescence (XRF) measurements.

Main Methods:

  • Experimental measurements of counting rates in grooved HPGe detectors below spectral peaks.
  • Postulation of charge collection anomalies and electric field distortions as the cause.

Main Results:

  • Observed time-variant counting rates in specific energy regions of grooved HPGe detectors.

Related Experiment Videos

  • Attributed these variations to prolonged or incomplete charge collection near groove bottoms.
  • Identified electric field distortions in the groove vicinity as the underlying mechanism.
  • Conclusions:

    • Grooved planar HPGe detectors exhibit count rate instability below spectral peaks due to charge collection issues.
    • This instability affects the minimum detectable count rate in XRF.
    • Collimating incident photons to the detector's central region can reduce these effects but decreases overall sensitivity.