Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Carbon dioxide rebreathing caused by deformed silicon leaflet in the expiratory unidirectional valve
Arumugam Vasudevan1, Satyen Parida, Lenin Babu Elakkumanan
1Department of Anaesthesiology and Critical Care, Jawaharlal Institute of Postgraduate Medical Education and Research, Puducherry, India.
Abstract:
Rebreathing of carbon dioxide caused by incompetent 'cage and disc' unidirectional valves has been reported earlier. Some manufacturers have changed the design of unidirectional valves to 'flexible leaflets'. We report a series of cases where a deformed membrane leaflet in expiratory unidirectional valves led to rebreathing of carbon dioxide.