Determination of lead and cadmium in seawater using a vibrating silver amalgam microwire electrode
Zhaoshun Bi1, Pascal Salaün, Constant M G van den Berg
1Department of Earth and Ocean Sciences, Liverpool University, Liverpool L69 3GP, UK.
Abstract:
Silver amalgamated electrodes are a good substrate to determine lead (Pb) and cadmium (Cd) in seawater because they have properties similar to mercury but without the free mercury (Hg). Here a silver amalgamated microwire (SAM) electrode is optimised for the determination of Pb and Cd in coastal waters and uncontaminated ocean waters. The SAM was vibrated during the deposition step to increase the sensitivity, and electroanalytical parameters were optimised. The Hg coating required plating from a relatively concentrated (millimolar) solution, much greater (500×) than used for instance to coat glassy carbon electrodes. However, the coating on the ex situ amalgamated electrode was found to be stable and could be used for up to a week to determine trace levels of Pb in seawater of natural pH. The limit of detection square-wave ASV (50 Hz) using the pre-plated SAM electrode was 8 pM Pb using a 1-min plating time at pH 4.5. The limit of detection in pH 2 seawater was 4 pM using a 5-min plating time, and it was 12 pM using a 10-min plating time at natural pH in the presence of air, using a square-wave frequency of 700 Hz. The vibrating SAM electrode was tested on the determination of Pb in reference seawater samples from the open Atlantic (at the 20 pM level), Pacific, and used for a study of Pb in samples collected over 24 h in Liverpool Bay (Irish Sea).
Related Concept Videos
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Electrodeposition
Electrodeposition can...
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...