Mixed-mode SPE for a multi-residue analysis of benzodiazepines in whole blood using rapid GC with negative-ion
Nerijus Karlonas1, Audrius Padarauskas, Arunas Ramanavicius
1Center of Nanotechnology and Material Science, Department of Analytical and Environmental Chemistry, Faculty of Chemistry, Vilnius University, Vilnius, Lithuania.
Abstract:
A sensitive and selective method for simultaneous quantitation of 15 benzodiazepines in human whole blood using rapid GC with negative-ion chemical ionization MS is proposed. A mixed-mode cation-exchange polymeric sorbent was used for SPE. Different extraction solvents or mixtures of solvents of different compositions for elution of the adsorbed analytes, and washing steps for eliminating interferences in the column were tested. Analytes were eluted from the column using 5% v/v NH4 OH in methanol. A derivatization step using different silylation reagents, time, and temperature was tested. Extracts from SPE were silylated by a mixture of N-(tert-butyldimethylsilyl)-N-methyltrifluoroacetamide, acetonitrile, ethyl acetate, and subjected to gas chromatographic analysis. The LODs of 15 benzodiazepines in whole blood samples ranged from 0.24-0.62 ng mL(-1). The RSDs of samples used for three different quality control concentration levels were lower than 7.0%, and the accuracy ranged from 89.5 to 110.5%. The results show that the developed method is accurate, sensitive, selective, and very fast. Finally, the applicability of this method for determination of trace concentrations of several benzodiazepines in real blood samples has been demonstrated.
More Related Videos
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
High-Performance Liquid Chromatography: Types of Detectors
Tandem Mass Spectrometry
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...