Integration of fluoroscopy-based guidance in orthopaedic trauma surgery - a prospective cohort study
Michael Kraus1, Stephanie von dem Berge, Hendrik Schöll
1Ulm University, Institute of Research in Rehabilitation Medicine, Wuhrstrasse 2/1, 88422 Bad Buchau, Germany; Federseeklinik Bad Buchau, Freihofgasse 14, 88422 Bad Buchau, Germany.
Introduction:
Computer-assisted guidance systems are not used frequently for musculoskeletal injuries unless there are potential advantages. We investigated a novel fluoroscopy-based image guidance system in orthopaedic trauma surgery.
Materials And Methods:
The study was a prospective, not randomised, single-centre case series at a level I trauma centre. A total of 45 patients with 46 injuries (foot 12, shoulder 10, long bones seven, hand and wrist seven, ankle seven and spine and pelvis four) were included. Different surgical procedures were examined following the basic principles of the Arbeitsgemeinschaft für Osteosynthesefragen/Association for the Study of Internal Fixation (AO/ASIF). Main outcome measurements were the number of trials for implant placement, total surgery time, usability via user questionnaire and system failure rate.
Results:
In all cases, the trajectory function was used, inserting a total of 56 guided implants. The system failed when used in pelvic and spinal injuries, resulting in a total failure rate of 6.5% (n=3) of all included cases. The overall usability was rated as good, scoring 84.3%.
Conclusion:
The novel image-guidance system could be integrated into the surgical workflow and was used successfully in orthopaedic trauma surgery. Expected advantages should be explored in randomised studies.