Dosimetric comparison between model 9011 and 6711 sources in prostate implants
1Department of Radiation Oncology, Indiana University School of Medicine, Indianapolis, IN 46202, USA. zhang248@iupui.edu
Comparing iodine-125 ((125)I) sources for prostate implants, this study found significant dosimetric differences between the 6711 and 9011 models. Using 6711 data for 9011 implants leads to inaccurate dose calculations, highlighting the need for source-specific data.
Area of Science:
- Medical Physics
- Radiation Oncology
- Nuclear Medicine
Background:
- Brachytherapy utilizes radioactive seeds for localized cancer treatment.
- Iodine-125 ((125)I) is a common isotope for prostate brachytherapy.
- Accurate dosimetry is crucial for effective treatment and minimizing side effects.
Purpose of the Study:
- To compare the dosimetric coverage of the 9011 iodine-125 ((125)I) source against the 6711 source in prostate implants.
- To evaluate potential errors in dose calculation when using the incorrect source data.
Main Methods:
- Postimplant dosimetry was performed on 18 prostate cancer patients (2 with 9011, 16 with 6711 sources).
- Implants were recalculated using the alternative source data within a commercial planning system.
- Dosimetric parameters (D100, D90, V200, V150, V100, V50, V20) were compared for prostate, bladder, rectum, and urethra.
Main Results:
- The 6711 source data overestimated coverage by 6.4% at D100 and 6.6% at D90 compared to actual 9011 data.
- Discrepancies up to 67% were observed at higher dose levels (V100, V150, V200), with average reductions of 2.7%, 14.6%, and 14.9% respectively.
- Significant differences were also noted in bladder, rectal, and urethral coverage.
Conclusions:
- There are clear dosimetric differences between the 9011 and 6711 iodine-125 ((125)I) sources.
- Utilizing 6711 source data for 9011 implants results in significant dose calculation errors.
- The 9011 source offers comparable dosimetric quality to the 6711 source when used correctly; 6711 data should not substitute for 9011 data.