Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 18, 2013
Normal and stenotic human aortic valve opening: in vitro assessment of orifice area changes with flow
J K Montarello1, A C Perakis, E Rosenthal
1Department of Cardiology, Guy's Hospital, London, U.K.
Abstract:
The ability to measure aortic valve area clinically has emphasized the need to understand the changes in aortic valve orifice area during flow. To compare the performance of normal and stenotic human aortic valves we used a pulsatile flow model that simulated in vivo flow conditions. Five normal autopsy specimens and 15 stenotic valves removed at operation were mounted into the model. Valve function was assessed by analysis of video recordings of valve leaflet motion during flow. Over the flow rates tested normal valves demonstrated a linear increase in orifice area. There was no resistance to leaflet opening and valve closure was rapid. The majority of stenotic valves demonstrated an increase in orifice area at low flow rates. No valve showed any increase in maximal area beyond flow rates of 3 l min-1. Increased leaflet resistance of these abnormal valves resulted in notably slower opening and closing rates. In patients with a high cardiac output and severe stenosis, overestimation of the anatomic orifice area derived by the Gorlin equation can result. This is not related to variability in maximal orifice area.
Related Concept Videos
Blood Flow
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....
Mitral Stenosis I: Introduction
Aortic Regurgitation I: Introduction
Pipe Flowrate Measurement
The orifice meter is a simple,...