Related Experiment Video
Updated: May 13, 2026

Augmented Reality Navigation-Guided Core Decompression for Osteonecrosis of Femoral Head
Published on: April 12, 2022
Evaluation of the accuracy of femoral component orientation by the CT-based fluoro-matched navigation system
Shinya Hayashi1, Takayuki Nishiyama, Takaaki Fujishiro
1Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, 7-5-1 Kusunoki-cho, Chuo-ku, Kobe, 650-0017, Japan. shayashi@med.kobe-u.ac.jp
Purpose:
Accurate orientation of acetabular and femoral components are important during THA. However, no study has assessed the use of the CT-based fluoro-matched navigation system during THA. Therefore, we have evaluated the accuracy of stem orientation by CT-based fluoro-matched navigation.
Methods:
The accuracy of stem orientation by CT-based fluoro-matched navigation was assessed by postoperative CT data. Furthermore, we compared the postoperative stem orientation with the intraoperative registration errors.
Results:
The average antetorsion error of the stem (navigation records - postoperative CT) was -0.5° ± 5.2°. The stem valgus error was 0.4° ± 2.7°. The accuracy of the navigation record for the orientation of the stem valgus was dependent on the intraoperative registration errors.
Conclusions:
The clinical accuracy of CT-based fluoro-matched navigation is adequate for stem alignment orientation, and the intraoperative verification of registration errors is valuable for checking the accuracy of stem orientation by navigation.