Test retest variability of TonoPen AVIA
Shibal Bhartiya1, Shveta J Bali, Mathew James
1Department of Ophthalmology, Glaucoma Facility, Dr. Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences, New Delhi, India, Vision Eye Institute, Chatswood, Australia, .
The Tonopen AVIA demonstrates excellent intra-session repeatability for measuring intraocular pressure (IOP) in both glaucoma patients and healthy individuals. This device offers reliable IOP readings for clinical use.
Area of Science:
- Ophthalmology
- Medical Devices
- Glaucoma Research
Background:
- Accurate intraocular pressure (IOP) measurement is crucial for glaucoma diagnosis and management.
- The Tonopen AVIA is a widely used handheld tonometer, but its intra-session repeatability requires robust evaluation.
Purpose of the Study:
- To assess the intra-session repeatability of IOP measurements using the Tonopen AVIA.
- To compare the repeatability in glaucoma patients versus healthy controls.
Main Methods:
- An observational study involving 180 eyes from 180 patients (50 with glaucoma, 130 controls).
- Three consecutive IOP measurements were taken using the Tonopen AVIA.
- Intraclass correlation coefficient (ICC) and coefficient of variation were calculated.
Main Results:
- The Tonopen AVIA showed high intra-session repeatability with ICC values ranging from 0.958 to 0.996.
- Coefficients of variation were 3.47% for glaucoma patients and 8.10% for controls, with an overall 6.07%.
- Coefficients of repeatability were 2.96 mmHg (glaucoma), 3.35 mmHg (controls), and 3.24 mmHg (overall).
Conclusions:
- The Tonopen AVIA exhibits good intra-session repeatability for IOP measurements.
- This repeatability is consistent across both glaucomatous and healthy eyes, supporting its clinical utility.
Related Concept Videos
Comparing Experimental Results: Student's t-Test
Significance Testing: Overview
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for: