Test retest variability of TonoPen AVIA
Related Experiment Video
Updated: May 13, 2026

Testing Tactile Masking between the Forearms
Published on: February 10, 2016
Shibal Bhartiya1, Shveta J Bali, Mathew James
1Department of Ophthalmology, Glaucoma Facility, Dr. Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences, New Delhi, India, Vision Eye Institute, Chatswood, Australia, .
Abstract:
The study was conducted to evaluate the intra-session repeatability of Tonopen AVIA (TPA). 180 eyes of 180 patients (50 eyes with glaucoma, 130 eyes of controls) were recruited for this observational study. The mean age of patients enrolled in the study was 43.9 ± 16.7 yrs (84 males, 96 females). Mean IOP recorded with Tonopen AVIA was 19.5 ± 9.5 mmHg, 19.4 ± 9.6 mmHg and 19.3 ± 9.2 mmHg, respectively in the first, second and third instances (P = 0.656). The intraclass correlation coefficient (ICC) ranged from 0.996 (95% CI: 0.956 - 0.998) for glaucoma subjects to 0.958 (95% CI: 0.934 - 0.975) for controls. The coefficient of variation in the study population ranged from 3.47% (glaucoma patients) to 8.10% (healthy controls), being 6.07% overall. The coefficient of repeatability varied between 2.96 (glaucoma patients), 3.35 (healthy controls) to 3.24 (overall). Thus, the Tonopen Avia shows good intrasessional repeatability of IOP in both glaucomatous patients and healthy subjects.
Related Concept Videos
Comparing Experimental Results: Student's t-Test
Significance Testing: Overview
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for: