In situ osmometry: validation and effect of sample collection technique
Related Experiment Video
Updated: May 13, 2026

Dissolved Solute Sampling Across an Oxic-Anoxic Soil-Water Interface Using Microdialysis Profilers
Published on: March 24, 2023
Moneisha Gokhale1, Ulrike Stahl, Isabelle Jalbert
1School of Optometry and Vision Science, The University of New South Wales, Sydney, New South Wales, Australia. m.gokhale@unsw.edu.au
Purpose:
To compare tear film osmolarity measurements between in situ and vapor pressure osmometers. Repeatability of in situ measurements and the effect of sample collection techniques on tear film osmolarity were also evaluated.
Methods:
Osmolarity was measured in one randomly determined eye of 52 healthy participants using the in situ (TearLab Corporation, San Diego, CA) and the vapor pressure (Vapro 5520; Wescor, Inc., Logan, UT) osmometers. In a subset of 20 participants, tear osmolarity was measured twice on-eye with the in situ osmometer and was additionally determined on a sample of nonstimulated collected tears (3 μL) with both instruments.
Results:
Mean (SD) tear film osmolarity with the in situ osmometer was 299.2 (10.3) mOsmol/L compared with 298.4 (10) mmol/kg with the vapor pressure osmometer, which correlated moderately (r = 0.5, P < 0.05). Limits of agreement between the two instruments were -19.7 to +20.5 mOsmol/L. Using collected tears, measurements with the vapor pressure osmometer were marginally higher (mean [SD], 303.0 [11.0] vs 299.3 [8.0] mOsmol/L; P > 0.05) but correlated well with those using the in situ osmometer (r = 0.9, P < 0.05). The mean (SD) osmolarity of on-eye tears was 5.0 (6.6) mOsmol/L higher than that of collected tears, when both measurements were conducted with the in situ osmometer. This was a consistent effect because the measurements correlated well (r = 0.65, P < 0.05).The in situ osmometer showed good repeatability with a coefficient of repeatability of 9.4 mOsmol/L (r = 0.8, P < 0.05).
Conclusions:
Correlation between the two instruments was better when compared on collected tear samples. Tear film osmolarity measurement is influenced by the sample collection technique with the osmolarity of on-eye tears being higher than that of collected tears. This highlights the importance of measuring tear film osmolarity directly on-eye. The in situ osmometer has good repeatability for conducting this measurement.
Related Concept Videos
Methods for Studying Drug Absorption: In situ
The Doluisio method involves perfusing a prepared segment of a rat's small intestine with a solution of radiolabeled drug and a non-absorbable marker. This helps to differentiate between absorbed and non-absorbed drug concentrations. The intestinal segment is connected at both ends using tubing and syringes,...
Osmosis and Osmotic Pressure of Solutions
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...