The REMEDEE trial: a randomized comparison of a combination sirolimus-eluting endothelial progenitor cell capture

Michael Haude1, Stephen W L Lee, Stephen G Worthley

  • 1Städtische Kliniken Neuss, Lukaskrankenhaus GmbH, Neuss, Germany. mhaude@lukasneuss.de

Insights

The Combo stent demonstrated noninferiority to the Taxus Liberté stent (PES) in reducing late lumen loss, showing comparable safety and efficacy in coronary artery stenting. Both drug-eluting stents had low rates of major adverse cardiac events and no reported stent thrombosis.

Area of Science:

  • Cardiovascular Medicine
  • Interventional Cardiology
  • Biomedical Engineering

Background:

  • Drug-eluting stents (DES) reduce restenosis and reintervention but are associated with late stent thrombosis and neoatherosclerosis.
  • Novel stent technologies are needed to overcome the limitations of current DES.
  • The sirolimus-eluting Combo stent with CD34 antibody coating presents an alternative approach.

Purpose of the Study:

  • To compare the efficacy and safety of the sirolimus-eluting Combo stent versus the paclitaxel-eluting Taxus Liberté stent (PES).
  • To evaluate angiographic, intravascular ultrasound, and clinical outcomes up to 12 months in the REMEDEE trial.
  • To assess the noninferiority of the Combo stent regarding in-stent late lumen loss.

Main Methods:

  • A randomized, controlled, multicenter trial (REMEDEE) involving 183 patients with de novo coronary artery stenoses.
  • Patients were randomized 2:1 to receive either the Combo stent or the PES.
  • Primary endpoint: in-stent late lumen loss at 9 months; Secondary endpoints: major adverse cardiac events (MACE).

Main Results:

  • The Combo stent was noninferior to PES for 9-month in-stent late lumen loss (0.39 ± 0.45 mm vs. 0.44 ± 0.56 mm; pnoninferiority = 0.0012).
  • At 12 months, MACE occurred in 8.9% of Combo patients and 10.2% of PES patients (p = 0.80), with no significant differences in mortality, myocardial infarction, or target lesion revascularization.
  • No stent thrombosis was reported in either stent group.

Conclusions:

  • The Combo stent met its primary noninferiority angiographic endpoint, demonstrating effectiveness.
  • The Combo stent proved safe, with a low rate of clinical events comparable to PES.
  • Both stent types showed no instances of stent thrombosis up to 12 months, suggesting a favorable safety profile.