Evaluation of a novel fascial dilator modified with scale marker in percutaneous nephrolithotomy for reducing the
Guohua Zeng1, Zhijian Zhao, Wen Zhong
11 Department of Urology, Minimally Invasive Surgery Center, The First Affiliated Hospital of Guangzhou Medical University , Guangzhou, China .
Objective:
To evaluate the role of the fascial dilators modified with a scale marker in reducing fluoroscopy time during percutaneous nephrolithotomy (PCNL).
Method:
In a randomized clinical trial, eligible 100 consecutive patients anticipated to undergo single-tract PCNL were randomly assigned into two groups by the closed envelope method, to compare the efficacy and safety of the modified scale fascial dilator (group 1, n=50) and traditional nonscale fascial dilator (group 2, n=50) with respect to the X-ray exposure duration as well as the outcome of PCNL.
Results:
There was no significant difference between group 1 and group 2 regarding to the initial successful access rate (100% v 96%), operative time (79.4 v 83.7 minutes), the initial stone-free rate (82% v 81.3%), complications (24% v 20%), and bleeding required blood transfusion (4% v 6%) (all p>0.05). The mean operative duration and fluoroscopic time for tract dilation in group 1 and group 2 were 164.4 ± 19 seconds v 168.3 ± 14 seconds, and 3.2 ± 0.9 seconds v 22.2 ± 4.6 seconds, (p=0.250 and p<0.001), respectively.
Conclusions:
Compared to the use of nonscaled dilators, using the scaled fascial dilators for a percutaneous access in PCNL was found to be associated with less radiation hazards along with similar stone-free rate and complications.