Filippo Randelli1, Lorenzo Banci, Sara Favilla
15th Orthopaedic Department (Department of Hip Surgery), Policlinico San Donato, Milan, Italy.
Abstract:
Patients with ASR implants (resurfacing and large-diameter (XL) metal-on-metal (MoM) total hip arthroplasty), even if asymptomatic and with a stable prosthesis, may present extremely high blood metal ion levels. We report on a consecutive series of fourteen ASR revisions, focusing on osteolysis and their radiographic correspondence and their correlation with blood metal ion levels. At revision, seven hips revealed severe periacetabular osteolysis which was radiographically undetectable in six and asymptomatic in five. Seven hips with no acetabular osteolysis had significantly lower serum Cr and Co ion concentrations (respectively 25.2, 41.1 μg/l) compared to the seven hips with severe acetabular bone loss (respectively 70.1, 147.0 μg/l). Elevated blood metal ion levels should be considered as a warning of undetectable and ongoing periprosthetic osteolysis in asymptomatic patients with ASR prosthesis.