The optimum Ga-67-citrate gamma camera imaging quality factors as first calculated and shown by the Taguchi's
Da Ming Yeh1, Pai Jung Chang, Lung Kwang Pan
1Graduate Institute of Radiological Science, Central Taiwan University of Science and Technology, Takun, TaiChung 406, Taiwan, R.O.C.
Abstract:
In this work gallium-67 ((67)Ga) gamma camera imaging quality was optimized using the Taguchi's analysis and a planar phantom. The acrylic planar phantom was LASER-cut to form groups of slits 1mm wide and 5mm deep, to determine the spatial resolution and contrast ratio that could be achieved in a (67)Ga citrate nuclear medicine examination. The (67)Ga-citrate solution was injected into the slits to form an active radioactive line source which was placed between regular acrylic plates for optimization. Then, nine combinations of four operating factors: L9 (3((4)), of the gamma camera imaging system were used and followed the Taguchi's analysis. The four operating factors were: a) the type of collimator in front of the NaI(Tl) detector, b) the region of interest of (67)Ga gamma rays spectrum, c) the scanning speed of NaI(Tl) detector head and d) the activity of (67)Ga. The original judged grade of the planar phantom image quality was increased 36% and factors a) and b) were confirmed to dominate. The cross interaction among factors was also discussed. Our results showed that the optimal factor settings of the gamma camera imaging system were verified by performing a routine nuclear medicine examination in ten cases. Nine cases showed the same optimal settings as estimated by three highly trained radio-diagnostic physicians. Additionally, the optimal setting yielded clearer images with greater contrast than did the conventional settings. In conclusion, this work suggests for practical use an optimized process for determining both the spatial resolution and the contrast ratio of a gamma camera imaging system using Taguchi's optimal analysis and a planar phantom. The Taguchi's method is most effective in targeting a single quality characteristic but can also be extended to satisfy multiple requirements under specific conditions by revising the definition of signal to noise ratio.

![Automated Preparation of [68Ga]Ga-3BP-3940 on a Synthesis Module for PET Imaging of the Tumor Microenvironment](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F68356.jpg&w=3840&q=50)