Clinical evaluation of the FreeStyle Precision Pro system
Ronald Brazg1, Kristen Hughes, Pamela Martin
1Rainier Clinical Research Center, Renton, WA, United States.
Background:
A new version of international standard (ISO 15197) and CLSI Guideline (POCT12) with more stringent accuracy criteria are near publication. We evaluated the glucose test performance of the FreeStyle Precision Pro system, a new blood glucose monitoring system (BGMS) designed to enhance accuracy for point-of-care testing (POCT).
Methods:
Precision, interference and system accuracy with 503 blood samples from capillary, venous and arterial sources were evaluated in a multicenter study. Study results were analyzed and presented in accordance with the specifications and recommendations of the final draft ISO 15197 and the new POCT12.
Results:
The FreeStyle Precision Pro system demonstrated acceptable precision (CV <5%), no interference across a hematocrit range of 15-65%, and, except for xylose, no interference from 24 of 25 potentially interfering substances. It also met all accuracy criteria specified in the final draft ISO 15197 and POCT12, with 97.3-98.9% of the individual results of various blood sample types agreeing within ±12 mg/dl of the laboratory analyzer values at glucose concentrations <100mg/dl and within ±12.5% of the laboratory analyzer values at glucose concentrations ≥100 mg/dl.
Conclusions:
The FreeStyle Precision Pro system met the tighter accuracy requirements, providing a means for enhancing accuracy for point-of-care blood glucose monitoring.
Related Concept Videos
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Pulse Assessment Sites
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):