A Bedside, Single Burr Hole Approach to Multimodality Monitoring in Severe Brain Injury
Published on: March 26, 2019
Zero drift of intraventricular and subdural intracranial pressure monitoring systems
Li Chen1, Hang-gen Du, Li-chun Yin
1Neurosurgical Department, Second Affiliated Hospital of Zhejiang Chinese Medical University, Hangzhou 310003, China.
Objective:
To assess zero drift of intraventricular and subdural intracranial pressure (ICP) monitoring systems.
Methods:
A prospective study was conducted in patients who received Codman ICP monitoring in the neurosurgical department from January 2010 to December 2011. According to the location of sensors, the patients were categorized into two groups: intraventricular group and subdural group. Zero drift between the two groups and its association with the duration of ICP monitor were analyzed.
Results:
Totally, 22 patients undergoing intraventricular ICP monitoring and 27 receiving subdural ICP monitoring were enrolled. There was no significant difference in duration of ICP monitoring, zero drift value and its absolute value between intraventricular and subdural groups (5.38 d+/-2.58 d vs 4.58 d+/-2.24 d, 0.77 mm Hg+/-2.18 mm Hg vs 1.03 mm Hg+/-2.06 mm Hg, 1.68 mm Hg+/-1.55 mm Hg vs 1.70 mm Hg+/-1.53 mm Hg, respectively; all P larger than 0.05). Absolute value of zero drift in both groups significantly rose with the increased duration of ICP monitoring (P less than 0.05) while zero drift value did not. Moreover, daily absolute value in the intraventricular group was significantly smaller than that in the subdural group (0.27 mm Hg+/-0.32 mm Hg vs 0.29 mm Hg+/-0.18 mm Hg, P less than 0.05).
Conclusion:
This study demonstrates that absolute value of zero drift significantly correlates with duration of both intraventricular and subdural ICP monitoring. Due to the smaller daily absolute value, ICP values recorded from intraventricular system may be more reliable than those from subdural system.
Related Concept Videos
Increased Intracranial Pressure l: Introduction
Increased Intracranial Pressure ll: Pathophysiology
Measurement of Blood Pressure
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Errors occurring during blood pressure monitoring
Several factors...