Circumferential rupture and entrapment of a balloon-on-a-wire device during coronary angioplasty

E Nukta1, B Meier, P Urban

  • 1Cardiology Center, University Hospital, Geneva, Switzerland.

Insights

During coronary artery interventions, a balloon-on-a-wire catheter ruptured, temporarily trapping it. The blockage was successfully treated using an alternative over-the-wire balloon catheter.

Area of Science:

  • Interventional Cardiology
  • Vascular Surgery

Background:

  • Percutaneous coronary intervention (PCI) is a common procedure for treating coronary artery disease.
  • Balloon angioplasty is a key technique used in PCI to dilate stenotic lesions.

Observation:

  • Circumferential rupture of a balloon-on-a-wire catheter occurred during dilation of a distal right coronary artery lesion.
  • The ruptured balloon became temporarily entrapped within the arterial stenosis.

Findings:

  • Successful retrieval of the entrapped balloon was achieved.
  • Completion of the lesion dilation was successfully performed using an over-the-wire balloon catheter.

Implications:

  • Highlights the potential risk of balloon catheter failure during complex PCI procedures.
  • Demonstrates the utility of alternative devices, such as over-the-wire balloon catheters, for successful lesion treatment after device malfunction.
  • Underscores the importance of device selection and preparedness for managing intraprocedural complications in interventional cardiology.