Inaccuracy in determining mean arterial pressure with oscillometric blood pressure techniques
Jorn Vos1, Hieronymus H Vincent, Marianne C Verhaar
1Department of Internal Medicine, St. Antonius Hospital Nieuwegein, Nieuwegein, The Netherlands.
Objective:
Accurate determination of MAP is important in the calibration of pressure waveforms for calculating central blood pressure (BP). Currently, a precise, individualized measurement of mean arterial pressure (MAP) can be obtained only with intra-arterial measurements of BP or with applanation tonometry. We conducted a study of whether easy-to-use oscillometric devices, validated for systolic and diastolic BP measurements (BHS protocol), give accurate determinations of MAP.
Methods:
We compared measurements of MAP made with the WatchBP Office oscillometric monitor in 102 subjects with values of MAP assessed by pulse-wave analysis (PWA) (SphygmoCor).
Results:
The mean (± SD) oscillometric MAP assessed with the WatchBP Office monitor was 97 ± 12.5 mm Hg, which was equivalent to 23.6 ± 9.1% of the pulse pressure (PP) above diastolic blood pressure (DBP). The MAP as assessed through PWA was 106 ± 14.6 mm Hg (P < 0.01), or 37.7 ± 3.9% of the PP above DBP. In simultaneous measurements made on both arms with the WatchBP Office monitor we observed individual differences in pressure in the left vs. the right arm.
Conclusions:
The MAP displayed by the WatchBP Office monitor is too imprecise to be used for calibrations. We suggest that devices for measuring BP not display MAP unless their accuracy is tested.
More Related Videos
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...