Related Experiment Video
Updated: May 12, 2026

Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
Published on: June 6, 2018
Analysis of 85Kr: a comparison at the 10 -14 level using micro-liter samples
1Hefei National Laboratory for Physical Sciences at Microscale, University of Science and Technology of China, Hefei, Anhui, China.
Abstract:
The isotopic abundance of (85)Kr in the atmosphere, currently at the level of 10(-11), has increased by orders of magnitude since the dawn of nuclear age. With a half-life of 10.76 years, (85)Kr is of great interest as tracers for environmental samples such as air, groundwater and ice. Atom Trap Trace Analysis (ATTA) is an emerging method for the analysis of rare krypton isotopes at isotopic abundance levels as low as 10(-14) using krypton gas samples of a few micro-liters. Both the reliability and reproducibility of the method are examined in the present study by an inter-comparison among different instruments. The (85)Kr/Kr ratios of 12 samples, in the range of 10(-13) to 10(-10), are measured independently in three laboratories: a low-level counting laboratory in Bern, Switzerland, and two ATTA laboratories, one in Hefei, China, and another in Argonne, USA. The results are in agreement at the precision level of 5%.
More Related Videos
14:53In Situ Detection and Single Cell Quantification of Metal Oxide Nanoparticles Using Nuclear Microprobe Analysis
Published on: February 3, 2018
12:48Measuring Cation Transport by Na,K- and H,K-ATPase in Xenopus Oocytes by Atomic Absorption Spectrophotometry: An Alternative to Radioisotope Assays
Published on: February 19, 2013