Related Experiment Video
Updated: May 12, 2026

07:57
Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Clinical application of gamma knife dose verification method in multiple brain tumors : modified variable ellipsoid
Beong Ik Hur1, Jae Min Lee, Won Ho Cho
1Department of Neurosurgery, School of Medicine, Pusan National University, Busan, Korea.
Journal of Korean Neurosurgical Society
|April 6, 2013
Summary
The modified variable ellipsoid modeling technique (MVEMT) accurately verifies Leksell Gamma Knife® (LGK) plans for multiple brain tumor targets. This independent method provides statistically identical data to the Leksell GammaPlan® (LGP) system, ensuring patient safety.
Area of Science:
- Medical Physics
- Radiation Oncology
- Neurosurgery
Background:
- Leksell Gamma Knife® (LGK) utilizes a single-fraction high-dose strategy, necessitating precise treatment planning.
- Independent verification of Leksell GammaPlan® (LGP) is crucial for patient safety and minimizing treatment errors.
- Existing verification methods lack statistical validation for multiple LGK target treatments.
Purpose of the Study:
- To evaluate the accuracy and statistical agreement of the modified variable ellipsoid modeling technique (MVEMT) for verifying LGP in multiple target treatments.
- To establish MVEMT as a reliable independent tool for LGP verification in radiosurgery.
Main Methods:
- A total of 500 points across 10 patients with multiple brain tumor targets were analyzed.
- Comparison between LGP and MVEMT data included dose at random points, maximal dose points, and target volumes.
- Statistical analysis employed t-tests and Bland-Altman plots to compare the two measurement techniques.
Main Results:
- No statistically significant difference in dose was found at 500 random points between LGP and MVEMT.
- Maximal dose differences between the methods ranged from -2.4% to 6.1%.
- An average distance of 1.6 mm was observed between maximal dose points when comparing LGP and MVEMT.
Conclusions:
- MVEMT demonstrated excellent agreement with LGP for radiosurgery planning involving multiple targets.
- MVEMT serves as a valuable, independent tool for multiple target treatment planning, yielding statistically identical results to LGP.
- MVEMT is suitable for use as a reference dose verification system for multiple brain tumors.