Effects of automated kilovoltage selection technology on contrast-enhanced pediatric CT and CT angiography
Marilyn J Siegel1, Charles Hildebolt, David Bradley
1Mallinckrodt Institute of Radiology, Washington University School of Medicine, 510 S Kingshighway Blvd, St Louis, MO 63110, USA. siegelm@mir.wustl.edu
Insights
Automated kilovoltage selection in pediatric CT scans significantly reduces radiation dose by 27% compared to standard protocols. This technology maintains diagnostic image quality and contrast-to-noise ratio (CNR) in children undergoing CT and CT angiography.
Area of Science:
- Radiology
- Medical Imaging
- Pediatric Imaging
Background:
- Computed tomography (CT) is crucial for pediatric diagnosis.
- Optimizing radiation dose in pediatric CT is essential.
- Automated kilovoltage selection offers potential for dose reduction.
Purpose of the Study:
- To evaluate the impact of automated kilovoltage selection on radiation dose.
- To assess changes in tube current and image quality in pediatric thoracoabdominal CT and CT angiography.
- To compare automated kilovoltage protocols with standard 120-kVp protocols.
Main Methods:
- Retrospective study of 87 pediatric patients (1 day to 19 years).
- Comparison of volumetric CT dose index (CTDIvol) and milliamperage between automated kilovoltage and standard 120-kVp protocols.
- Subjective image quality scoring and quantitative measurement of contrast, noise, and contrast-to-noise ratio (CNR) in a subset of patients.
Main Results:
- Automated kilovoltage reduced median CTDIvol by 27% (4.8 mGy vs. 7.1 mGy).
- Kilovoltage was reduced from 120 kVp in 94% of patients, with greater dose reduction in CT angiography.
- Diagnostic image quality was maintained, with similar CNRs for liver and portal vein despite increased contrast and noise.
Conclusions:
- Automated kilovoltage selection significantly lowers radiation dose in pediatric CT and CT angiography.
- This technology effectively preserves diagnostic image quality and CNR.
- Automated kilovoltage is a valuable tool for dose optimization in pediatric imaging.
Purpose:
To assess effects of an automated kilovoltage selection tool on radiation dose, tube current, and image quality in pediatric contrast material-enhanced thoracoabdominal computed tomography (CT) and CT angiography.
Materials And Methods:
The institutional review board approved this retrospective study of 87 children (median age, 10.5 years; age range, 1 day to 19 years) who underwent thoracic CT, abdominal CT, or CT angiography with an automated kilovoltage protocol between November 2010 and July 2011. Volumetric CT dose index (CTDIvol) and milliamperage were compared with estimated prescanning calculations from a standard 120-kVp protocol. Image quality was subjectively scored on a scale of 1 to 5. In 15 patients who underwent scanning with both automated kilovoltage and standard 120-kVp protocols, image contrast and noise were quantitatively measured and the contrast-to-noise ratio (CNR) was calculated. Parametric and/or nonparametric testing of independent and/or paired samples was performed.
Results:
Estimated median CTDIvol for all 87 studies was 4.8 mGy for the automated kilovoltage protocol versus 7.1 mGy for the 120-kVp protocol, yielding a 27% dose reduction (25% and 75% percentiles: 23% and 37%, respectively; P < .001). The median milliamperage was 125.0 with the automated protocol and 87 with the 120-kVp protocol (P < .001). Use of automated kilovoltage selection resulted in kilovoltage reduction from 120 kVp in 82 of the 87 patients (94%), with the percentage dose reduction greater for CT angiography than for chest CT (P < .001). Reduction in kilovoltage showed a significant correlation with patient size (P < .01). Diagnostic image quality was achieved in all patients. Mean contrast and noise were higher with the automated kilovoltage protocol than with the 120-kVp protocol (P ≤ .04), but CNRs for liver and portal vein were similar (P = .54 and .73, respectively).
Conclusion:
In pediatric CT and CT angiography, the use of automated kilovoltage selection technology significantly reduces radiation dose while maintaining image quality and CNR.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies VII: Vascular Imaging