Radiation exposure in multidetector CT: dose comparison between late 1990s and early 2010s in Korea
Moon Chan Kim1, Dong Kyoon Han, Yon Min Kim
1Department of Radiologic Science, Eulji University, Kyungki, Korea.
Abstract:
The effective dose under ordinary clinical computed tomography (CT) protocols using three kinds of 64-channel and a 40-channel CT ranged from 0.6 to 15.5 mSv in early 2010s. And the organ dose ranged from 1.6 to 130.4 mGy: orbital and brain doses for brain stroke CT were the highest. For the comparison of the effective dose between late 1990s and early 2010s, multidetector CT (MDCT) for high-resolution lung CT was 2.4 times higher than that of single-detector CT (SDCT) and the ratio was the highest. However, the effective dose at MDCT was 20 % lower than that of SDCT in chest CT due to applying dose-saving techniques. In organ dose comparisons, high-resolution lung CT at MDCT was 3.5-4.5 times higher than that of SDCT, and showed 1.1-1.5 times higher than that of SDCT in the head and chest CT. For the abdomen CT, the primary organ dose at MDCT was ∼30 % lower than that of SDCT.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies for Cardiovascular System V: CT
Biological Effects of Radiation
X-ray Imaging